PAT700 Total Organic Carbon Analyzer – Bottle Mode
Page 75
Anatel
Operator Manual
Chapter 6
Bottle Mode
6.1
Calibration and validation
Bottle mode allows you to perform calibration, validation, and system suitability test
procedures on the PAT700. Proper calibration of the analyzer is critical to maintaining
optimum instrument performance. You may perform such functions automatically with the
Onboard Automated Standards Introduction System (OASIS™) using the touch-screen
interface. This method minimizes user intervention by using a prepackaged standards kit with
RFID technology. You may run calibration, validation, and system suitability test procedures
immediately or schedule them to run later.
This chapter provides
general
instructions on performing calibration and validation
procedures. Refer to the specific Standard Operating Procedure (SOP) for detailed
instructions.
•
Calibration assures the accuracy of instrument’s readings at regular intervals.The
PAT700 is calibrated at the factory before shipment. You do not need to perform a
TOC calibration before placing the analyzer into service.
•
Validation ensures the accuracy of the calibration by challenging the calibration with
the analysis of a known standard.
•
Calibration and validation involve running analyses of one or more standards and
blanks (protocols) to determine analyzer response.
•
In bottle mode, the analog output retains the last online TOC, conductivity, and
temperature values, and TOC and conductivity alarms are deactivated.
Four functions are available under bottle mode: run standards, grab sample, excursion mode,
and unload bottles.
Onboard Automated Standards Introduction System (OASIS)
The PAT700 Onboard Automated Standards Introduction System (OASIS) is used for
performing standards tests, grab sampling, and excursion sampling. Through the bottle mode,
various bottle test functions can be selected. The OASIS system is automated through the
use of RFID technology (see
). When standards bottles with RFID
tags are installed in the system, the analyzer uses RFID technology to read and transmit data
about the standards to the analyzer. This system eliminates the need for manual data entry.
Although equipped to perform automatically using RFID tagged standards, you may also use
bottles without RFID tags by manually entering the standards data when prompted by the
analyzer.
TOC calibration protocol
Select up to three levels (250, 500 and 750 ppb) of a sucrose standard. For maximum
accuracy, the procedure allows up to three analysis replicates per standard.
TOC validation protocol
Perform a single-point or multi-point validation of the TOC calibration using up to three levels
of sucrose validation standards.
Содержание ANATEL PAT700
Страница 1: ...ANATEL PAT700 ON LINE TOTAL ORGANIC CARBON ANALYZER Operator Manual...
Страница 2: ......
Страница 4: ......
Страница 56: ...Page 46 Run Modes PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 112: ...Page 102 Bottle Mode PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 136: ...Page 126 Data Review PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 154: ...Page 144 Specifications PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 156: ......
Страница 157: ......