Page 68
Setup – PAT700 Total Organic Carbon Analyzer
Operator Manual
Anatel
WG
M - 2
8
No
vemb
e
r 20
07
- Ed
it
io
n 0
1
Return to online mode checkbox
tells the analyzer whether or not to return to online mode
following the calibration. You may enable or disable (default) this option. When enabled, the
analyzer returns to online mode following the calibration. When disabled, the analyzer stays
at the bottle mode dialog box. All valves are closed, the pump is off, and the lamp is off.
Slope limit from previous cal text box
allows you to specify the acceptable percent
difference between the resulting gain factor and the current value. You may enter a value from
1 to 5%. The default value is 5%.
Slope limit from factory cal text box
allows you to specify the acceptable percent
difference between the resulting gain factor and the factory value. You may enter a value from
1 to 10%. The default value is 10%.
Use calibration resistor check box
allows the use of the calibration check resistor during
the conductivity calibration process. See
Conductivity calibration, page 85
.
Return icon
returns you to the setup dialog box and saves any changes. Any changes are
noted in the audit trail.
TOC validation tab
Figure 5-22 Bottle test setup dialog box, TOC validation tab
Return to online mode checkbox
enables the analyzer to return to the online mode
following the validation. You may enable or disable (default) this option. When enabled, the
analyzer returns to online mode following the validation. When disabled, the analyzer stays at
the bottle mode dialog box. All valves are closed, the pump is off, and the lamp is off.
Return icon
returns you to the setup dialog box and saves any changes. Any changes are
noted in the audit trail.
Содержание ANATEL PAT700
Страница 1: ...ANATEL PAT700 ON LINE TOTAL ORGANIC CARBON ANALYZER Operator Manual...
Страница 2: ......
Страница 4: ......
Страница 56: ...Page 46 Run Modes PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 112: ...Page 102 Bottle Mode PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 136: ...Page 126 Data Review PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 154: ...Page 144 Specifications PAT700 Total Organic Carbon Analyzer Operator Manual Anatel WGM 28 November 2007 Edition 01...
Страница 156: ......
Страница 157: ......