page 31 / 33
© HAAG-STREIT AG, CH-3098 Koeniz, Switzerland
1500.7001000.04050
Diagramm
Spektrale Bestrahlungsstärke E
λ
bei maximaler
Intensität und maximaler Spaltöffnung
Diagramme
Puissance spectrale E
λ
aux valeurs maxima-
les d'intensité et d'ouverture
Diagram
Spectral irradiance E
λ
at maximum intensity and
maximum aperture
7.5 V (Lampe / lampe / lamp)
Wellenlänge / Longueur d'onde / Wave length [nm]
Spektrale Bestrahlungsstärke / Puissance spectrale / Spec-
tral irradiance
[mW/(cm
2
nm)]
Appendix C
Technical specifications
Slit illumination
Slit image width
0 - 8 mm continuous
Slit image length
1 - 8 mm continuous
Illumination field diameter
8; 5; 3; 2; 1; 0.2 mm
Test mark with fixation star
Slit image radial range
±
90
°
Radial movement of the slit illumination relative to the
microscope axis
horizontal
±
90
°
, vertical 0 - 20
°
Filters
blue, redfree (green), grey (10 %),
heat absorption filter permanently
mounted
Light source
Tungsten incandescent
light bulb 6 V / 4.5 A
Illumination intensity
0 V / 0 Lux
at position 0
5 V / 150 000 Lux
at position 1/2
6 V / 300 000 Lux
at position 1
7.5 V / 600 000 Lux at position 2
Irradiance depending on light wave length
see diagram at right hand side
Irradiance at max. intensity (7.5 V lamp) and
max. aperture*)
0.07 mW / cm
2
(300 nm - 400 nm)
180 mW / cm
2
(380 nm - 700 nm)
100 mW / cm
2
(700 nm - 1100 nm)
Spectrally-weighted photochemical aphakic output source
radiance L
A
at maximum intensity (7.5 V lamp) and maximum
aperture: 450 mW / (cm
2
sr).*)
Spectrally-weighted photochemical phakic output source ra-
diance L
B
at maximum intensity (7.5 V lamp) and maximum
aperture: 430 mW / (cm
2
sr).*)
Spectrally-weighted photochemical radiance L
A
and L
B
give
a measure of the potential hazard that exists for a beam of
light to cause photochemical damage to the retina. L
A
gives
the measure either for eyes in which the crystalline lens
has been removed (aphakes) and has not been replaced by
a UV-blocking lens or for eyes of very young children. L
B
gi-
ves the measure for eyes in which a crystalline lens is in
place. Values of L
A
and L
B
over 80 mW / (cm
2
sr) are consi-
dered high for beams which completely fill a dilated pupil. *)
*)
According to ISO 10939
300
0
0.2
0.4
0.6
0.8
1
1.2
500
700
900
1100