page 23 / 33
© HAAG-STREIT AG, CH-3098 Koeniz, Switzerland
1500.7001000.04050
Appendix A Installation of the equipment
A.1
Mounting the microscope and the
illumination unit
Place the stereo microscope on the saddle bearing of the
microscope arm. Move it forward against the stop piece and
secure with the screw (13) appearing from obliquely below
(use the Allen key 5 mm).
Secure the breath shield (26) by means of the knurled screw
(27) on the inner side of the bearer arm.
Before attaching the illumination unit to the column of the
instrument base, the set screw (53) in the hub of the bearer
arm is to be sufficiently loosened so that its tip no longer
protrudes into the screw hole.
Grasp the illumination unit arm (52) as shown and place in
position.
Tighten the set screw (53) in the hub of the bearer arm with the
small screwdriver.
Small screwdriver
HS-Part number 1001436
Allen key 5 mm
HS-Part number 1001602
A.2
Mounting of headrest and power supply
The corresponding HAAG-STREIT operating instructions
should be observed!
13
27
26
52
53