optional settings
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 96 of 176
[6.B2]
Standard settings of administrator options
[6.B2]
Option Description
Standard
setting
Adjustable
Range
OP 01
Set back of error messages E 20, E 21,
E 30, E 31, E 32, E 41, E 50 and E 51
OP 02
OEM-Code
depending on the model
0 = h/p/cosmos (Standard)
1 = JAEGER / VIASYS /
Cardinal Health
2 = Ergo-Fit (1997-98)
3 = Proxomed / Kardiomed
4 = KISTLER Gaitway
6 = COSMED
9 = SCHILLER
OP 03
Selection of device type
- see list [6.B3] -
depending on the model
(depending on the model)
OP 04
Maximum speed for reverse belt rotation depending on the device type in
OP03
1.0 km/h .... max. speed
OP 05
Maximum Speed
depending on the device type in
OP03
1.0 km/h .... max. speed
OP06
Max. acceleration time in seconds
Increment 131 = 5 sec.
Increments: 255 ... 1
Seconds: 2.6 ... 655.4
OP 07
Maximum Elevation
depending on the device type in
OP03
0 .... max. elevation
OP 08
Heart Rate Display Interval
0 = beat to beat
0 or 1 ... 9 seconds
OP 09
Display Test for all displays
OP 10
Enter the serial number of the device
300…
12-digit serial number
OP 11
Activating and de-activating speed
measurement
depending on the device type in
OP03
0 = without speed measurement
1 = with speed measurement
OP 12
Setting of allowance for speed-signal-
comparison
(Measurement / Control)
6 %
This value does not reflect the
speed accuracy of the running-
belt.
0 = no comparison
6 = 6 % deviation allowed
8 = 8 % deviation allowed
10 = 10 % deviation allowed
OP13
Motor brake
OFF
OFF = not active
ON = active
OP 14
Maximum acceleration time for RS232
Increment 131 = 5 sec
Increments: 255 ... 1
Seconds: 2.6 ... 655.4
OP 16
Activating and de-activating heart rate
measurement
(depending on the model)
0 = HF measurement de-activated
1 = HF measurement activated,
HFU1 + RE06REC
2 = HF measurement activated,
HFUi;
auto = autom. detection
OP 33
Distance between rungs (only for ladder
ergometer)
254 = 25.4 cm
Distance between rungs in 0.1 cm
OP 34
Manual Speed Calibration
(depending on the device type;
OP 03)
Range from: 0 to 65535