operation
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 55 of 176
[5.B13]
2 Graded
test
Continuation
...
In an already pre-defined step-test it is possible, to set the pre-
programmed pause-times individually during the test, i.e. the
pause can be prolonged or ended immediately.
End pause manually:
Once the test profile has reached the pause mode and the
running belt has come to halt, you can select the next program
step immediately with START – either with warning sound
signals at the beginning by pressing START once or without by
pressing START twice. This is done independently of the pre-
defined pause-times and has no influence on the following run
of the program, i.e. the following program steps remain
unaffected. By pressing the START button the next speed
step scheduled is started. The pause can also be ended with
the „+“ button. In this case the speed set depends on the point
of release of the „+“ button.
Prolong pause:
Once the test profile has reached the pause mode and the
running belt has come to halt, it is possible to freeze pause
with the „-“ button.
=
indicates: „PAUS“
To end pause you can either press START or „+“ - see end
pause manually.
3 Conconi
test
e.g. for determination of anaerobic threshold via
heart rate curve of the subject
Endurance test (max. heart rate test)
Standard load profile:
N
Starting speed: 8.0 km/h, must be changed according the
conditions of the subject
N
Circuit (lap length): 200m (can be changed)
N
Increment: 0.5 km/h (can be changed)
STOP must be activated manually by the medical doctor when
the subject is fully exhausted.