technical data
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh [email protected] created 01.09.2008 printed 08.10.2008 page: 129 of 176
h/p/cosmos mercury
h/
p/
cosmos st
ra
to
s [a]
h/p/cosmos quas
ar
h/p/cosmos stellar [b]
h/p/cosmos mercury med
h/
p/
cosmos st
rat
os med [
c]
h/p/
co
sm
os
15
0/5
0 lo
com
otio
n D
E
me
d
h/p/
co
sm
os
15
0/5
0 lo
com
otio
n for
ce
DE m
ed
h/p/cosmos quas
ar med
h/p/cosmos stellar m
ed [d]
h/p/cosmos pulsar
h/p/cosmos pulsar 3p
Dimension of running surface
length x width
1500 x 500 mm
1700 x 650 mm
1500 x 500 mm
1700 x 650 mm
1900 x 650 mm
1900 x 650 mm
Dimension frame
length x width x height
(UserTe 150 mm width)
2100 x 800 x
1370 mm
2300 x 1050 x
1450 mm
2100 x 800 x
1370 mm
2300 x 1050 x
1450 mm
2500 x1050 x
1450 mm
2500 x1050 x
1450 mm
Track access height from floor
180 mm
230 mm
180 mm
230 mm
230 mm
230 mm
Weight of device
190 kg
[a] 180 kg
370 kg
[b] 360 kg
200 kg
[c] 190 kg
380 kg
[d] 370 kg
400 kg
440 kg
Weight load limits to floors
(traffic load to DIN 1055 part 3)
8.40 kN/m²
7.00 kN/m²
8.50 kN/m²
7.10 kN/m²
6.60 kN/m²
6.90 kN/m²
In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight
200 kg (deviation on request specially made)
Environmental conditions:
Transport and storage
temperature: –30...+50 °C
humidity: 0...95% - without condensation
barometric pressure: 700...1060hPa
Environmental conditions:
Operation
temperature: +10...+40 °C (deviation on request)
humidity: 30...70% - without condensation (up to 95% on request)
barometric pressure: 700...1060hPa
Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Central lubrication system
manual
Store program space
6 training profiles / 10 test profiles / 8 user definable profiles
RS232 Digital interface / serial port
(USB-converter optional)
COM1: standard; COM2: optional; COM3: optional (standard for service);
COM4: optional
COM1: standard; COM2: standard
COM3: optional (standard for
service); COM4: optional
Heart rate measurement and heart
rate dependent endurance control
1 channel receiver, ECG precisely
PC software
h/p/cosmos para control 4.0
®
included
options: h/p/cosmos para graphics
®
and h/p/cosmos para analysis
®
h/p/cosmos para control 4.0
®
and
h/p/cosmos para graphics
®
included
Reverse belt rotation / downhill
1
option
/
1
[a] and [b] not applicable
option /
1
[c] and [d] not applicable
standard
Safety Arch with harness***
option
Handrails
options: short, long, adjustable (only mercury and stratos), crossbar (standard with quasar, stellar and pulsar)
Arm support / additional keyboard
option
* Overload and weak power supply (undervoltage) could lead to higher discrepancies concerning the speed-display or dripping of the fuse. ** The time of
operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be
used again after short cool down period. Treadmills with a running surface 150/50 built approx. before June 2007 have a maximum elevation of 24%.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety harness with chest belt) is necessary and highly
recommended. When using a front-bar handrail (crossbar), the speed is limited to 5 km/h for safety reasons. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & EO.