22
.2 Graded test
22
.3 Conconi test
Description
Illustration
(e.g. for performance diagnostics based on lactate measurement)
Refer to “test mode” for safety information and adjustment.
Parameter
Default value
Starting speed
8.0 km/h
Increment
2.0 km/h
Acceleration level
4
Step length
3:00 min
Break time
0:30 min
Each parameter is adjustable.
STOP must be activated manually by the medical doctor.
Skip remaining break time:
Press “start” once
restart after countdown
Press “start” twice
restart immediately
Prolong break:
Press “-“ within break
“pause” is indicated
Press “start” to continue
Description
Illustration
(e.g. for performance diagnostics based on heart rate measurement)
Refer to “test mode” for safety information and adjustment.
Endurance test (max. heart rate test)
Standard load profile:
❚
Starting speed: 8.0 km/h, must be changed according to the condition
of the subject
❚
Circuit (lap length): 200 m (can be changed)
❚
Increment: 0.5 km/h (can be changed)
STOP must be activated manually by the medical doctor
❚
150/50 LC 150/50 150/50 new 170-190/65 200-300/75-125 emotion MCU6
Copyrigth: h/p/cosmos sports & medical gmbh
operation manual: cos105000_170-190_65_en rev. 1.05 dated 22.09.2021
page 72 of 97