.13 Software, Programs
Description
Data
para control
optional - external treadmill control
gaitway
optional - force measurement
Noraxon
optional - motion analysis
Microgate Optogait
optional - optical measurement of gait parameters
zebris Rehawalk
optional - pressure distribution measurement
Number of stored programmes
8 training profiles,
10 test profiles,
more than 100 user definable profiles
For other coscom
v4
compatible software see website
www.coscom.org
1
7
.14 All-pole disconnection
The following options are available for all-pole disconnection from voltage power supply:
❚
Unplug device from power socket.
❚
Unplug cable from device (if possible).
❚
Switch off device protection switch.
❚
Trigger residual current circuit breaker / RCD of the building.
Maintain enough free space to ensure access to cables, power plugs and the circuit breaker (see “position of subject and operator”).
For some RCDs type B (measuring also direct current) an additional leakage current limiter cos103344_170-65nf has to be installed
in treadmills, which do not have a potential isolation transformer.