![background image](http://html.mh-extra.com/html/h-p-cosmos/170-190-65-series/170-190-65-series_original-instructions-for-use_574520033.webp)
11
Operation
11
.1 General application procedure
Description
Illustration
Disinfect the device (see “cleaning”).
Disconnect the device and all accessories from mains
power supply before cleaning or disinfection.
Make sure…
…the PE-cable is connected to electrical installation
and device (if applicable),
…the device is directly plugged into the dedicated wall
socket, except if the treadmill is part of a ME system
including satellite PC med,
…the expulsion fuse on the front of the device is
switched on,
…all emergency stops are released.
Switch the running machine on with the on/off switch,
light goes on, system starts up (approx. 1 min).
Make sure the space under the treadmill is free from
persons, body parts or objects, especially when
switching on (treadmill will lower during initialization)
and when changing the elevation.
Normal condition: When starting, all displays show
“0”-values.
Select an operation mode.
For a detailed description, see following chapters.
Perform application.
-
Switch the running machine off with the on/off switch at
the terminal by pushing the button for more than 2 sec
(light goes off). By pushing the button under 2 sec., the
system goes in stand by modus.
Disinfect the device (see “cleaning”).
Disconnect the device and all accessories from mains
power supply before cleaning or disinfection.
❚
170-190/65 MCU6
Copyrigth: h/p/cosmos sports & medical gmbh
operation manual: cos105000_170-190_65_en rev. 1.05 dated 22.09.2021
page 33 of 97