GRAPH MODE
93
Set horizontal axis (Frequency)
Background
The X (horizontal) axis is selectable from Voltage
and Frequency sweep.
When Voltage sweep is selected: measurement
Frequency is fixed
When Frequency sweep is selected:
measurement Voltage is fixed
Select Frequency
1.
Press the Up/Down key and move
the cursor to Sweep.
Sweep:
Drive Level
2.
If necessary, press the Left/Right
key to change the sweep setting to
Frequency.
Drive Level
Frequency
Set start
Frequency
3.
Press the Up/Down key and move
the cursor to Start.
Start:
:
20.000Hz
Use the numerical keys to enter the
starting Frequency.
Range
20Hz ~ 1MHz/5MHz/10MHz
(stop Frequency must be higher than the start
Frequency)
50Hz
Enter
1.2kHz
k
Enter
Backspace
All clear
Clear
If a wrong unit is entered, the value is cancelled.
www.
.com
1.800.561.8187