LCR-8000G Series User Manual
78
Set frequency
Move the cursor to Freq column.
Enter the frequency using the
numerical keys and unit keys.
Func
Freq
500.00
Ls
2.00 V
Volt
Range
20Hz ~ 1MHz/5MHz/10MHz
5 digit resolution
Example:
0.5kHz (500Hz)
k
Enter
Set voltage
Move the cursor to Volt column.
Enter the voltage using the
numerical keys and unit keys.
Freq
Volt
2.00 V
500.00
Bias
Range
(1mV step)
10mV ~ 2V (DC or AC≤3 MHz)
10mV ~ 1V (AC>3 MHz)
Example:
100mV
m
V/A
Enter
Select data
capture rate
Move the cursor to Spd column.
Select the capture rate by
pressing F1 (Prog) repeatedly.
Bias
Spd
MAX
1.0000H
Hi
F 1
DC
AC≤
100Hz
AC≤
2kHz
AC>
2kHz
AC≥
1MHz
Slow
900ms 1.3s
600ms 600ms 620ms
Med
120ms 1.2s
470ms 450ms 470ms
Fast
60ms
650ms 180ms 150ms 150ms
Max
30ms
600ms 120ms 75ms