LCR-8000G Series User Manual
116
:meas:speed
<disc>
:meas:speed?
Selects or returns measurement speed.
Send parameter: max, fast, med, slow
Return parameter: 0 (max), 1 (fast), 2 (med), 3
(slow)
:meas:test:ac
Selects AC measurement.
:meas:test:rdc
Selects Rdc measurement.
:meas:test?
Returns measurement type.
Parameter: 0 (AC measurement), 1 (Rdc
measurement)
:meas:trig
Triggers an AC or Rdc measurement manually.
Returns the 1
st
and 2
nd
measurement (only the 1
st
in
Rdc).
Example: –396.283E-6, 99.558 (uF/D)
:mode?
Query the currently selected operating mode.
:rep <disc>
:rep?
Enables or returns repetitive measurements when
unit is returned to local control.
Send parameter: on (repetitive), off (single shot)
Return parameter: 0 (single shot), 1 (repetitive)
Example: :rep on (repetitive mode)
:trig
Triggers a measurement in the current mode.
Multi-step program command
:multi:set
Switches to the multi-step set-up page.
:multi:del
Removes a step in the program.
Parameter: 1 ~ 30
Example: :multi:del 2 (deletes step 2)