LCR-8000G Series User Manual
134
≦
10 kHz O/S Trimming Point List
The frequency of the extra trimming point is the measurement
frequency you were using before entering the calibration mode.
“O/S Trim Error!” warning will still appear after running a
≦
10
kHz trimming if you choose to use a measuring frequency higher
than 10 kHz or other than where you were when you ran the
≦
10
kHz trimming.
LCR Models
LCR-8101G
LCR-8105G
LCR-8110G
Trimming Point
7 points
7 points
7 points
1
20 Hz
20 Hz
20 Hz
2
100 Hz
100 Hz
100 Hz
3
300 Hz
300 Hz
300 Hz
4
1 kHz
1 kHz
1 kHz
5
3.3 kHz
3.3 kHz
3.3 kHz
6
10 kHz
10 kHz
10 kHz
7
Extra Trimming
Point
Extra Trimming
Point
Extra Trimming
Point
Spot Frequency O/S Trimming Point’s List
There is no extra trimming point if you choose to use Spot Freq. O/S
trimming.
“O/S Trim Error!” warning will still appear after running a spot
frequency trimming if you choose to use a measuring frequency
other than where you were when you ran the spot frequency O/S
trimming.
LCR Models
LCR-8101G
LCR-8105G
LCR-8110G
Trimming Point
1 point
1 point
1 point
1
The frequency of the sole trimming point is the
very measuring frequency you were using before
entering the calibration mode.
www.
.com
1.800.561.8187