GRAPH MODE
91
Horizontal Scale Setting
Set horizontal axis (Voltage)
Background
The X (horizontal) axis is selectable from Voltage
and Frequency sweep.
When Voltage sweep is selected: measurement
Frequency is fixed
When Frequency sweep is selected:
measurement Voltage is fixed
Select Drive level
(Voltage)
1.
Press the Up/Down key and move
the cursor to Sweep.
Sweep:
Frequency
2.
If necessary, press the Left/Right
key to change the sweep setting to
Voltage (Drive Level).
Frequency
Drive Level
Set start Voltage
3.
Press the Up/Down key and move
the cursor to Start.
Start:
:
50mV
Use the numerical keys to enter the
starting Voltage.
Range
10mV ~ 2V (AC≤3 MHz)
10mV ~ 1V (AC>3 MHz) *1mV step
100mV
m
V/A
Enter
1V
V/A
Enter
Backspace
All clear
Clear
If a wrong unit is entered, the value is cancelled.