LCR-8000G Series User Manual
110
Configure GPIB interface
Connection
Connect the GPIB cable to the
rear panel port: 24-pin female
connector.
Pin assignment
12
1
24
13
Pin1
Data line 1
Pin13
Data line 5
Pin2
Data line 2
Pin14
Data line 6
Pin3
Data line 3
Pin15
Data line 7
Pin4
Data line 4
Pin16
Data line 8
Pin5
EOI
Pin17
REN
Pin6
DAV
Pin18
Ground
Pin7
NRFD
Pin19
Ground
Pin8
NDAC
Pin20
Ground
Pin9
IFC
Pin21
Ground
Pin10
SRQ
Pin22
Ground
Pin11
ATN
Pin23
Ground
Pin12
Shield (screen)
Pin24
Signal ground
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
www.
.com
1.800.561.8187