PASS-FAIL MODE
61
Single-Step Test Configuration
Overview
Background / test
type
Pass/Fail test checks whether the measurement
result sits between the Hi(high) and Lo(low) limit.
Three methods are available: absolute limit,
percentage and delta limit.
Absolute
limit
The Hi and Lo limit are defined by
absolute values.
Lo
Limit
Hi
Limit
Pass
Percentage
limit
The Hi and Lo limit are defined by the
distance (percentage) from the
Nominal value.
Nominal
– ##%
Pass
Nominal
Nominal
+ ##%
Delta limit
The Hi and Lo limit are defined by the
difference from the nominal value.
Nominal
– value
Pass
Nominal
Nominal
+value
Test item
C
S
Series capacitance
X
Reactance
C
P
Parallel
capacitance
B
Susceptance
L
S
Series inductance
Z
Impedance
L
P
Parallel inductance
Y
Admittance
R
S
Series resistance
R
DC
DC Resistance