GETTING STARTED
25
Fixture connection
Panel operation
1.
Discharge the test component before
connecting the fixture set.
2.
Connect each fixture terminal to the front panel
BNC connector with matching color.
3.
Connect the fixture to the test component. If the
component has polarity, connect the H side to
the positive lead and the L side to the negative
lead. Make sure the distance between the lead
base and fixture clip is short enough.
H side
L side
4.
If the test component has an outer case
unconnected to either of the leads, connect the
ground terminal for noise level reduction.
www.
.com