ADVANCED
MEASUREMENT
41
Advanced Measurement Overview
Background
Advanced measurement mainly refers to the type of
measurement which uses the result obtained by one of
the basic measurements: ACV, DCV, ACI, DCI, 2/4W,
Diode/Continuity, Frequency/Period, and Temperature.
Advanced
Measurement
Basic Measurement
AC/DCV
AC/DCI
2/4W
Hz/P
°C/°F
dB
●
— — — — —
dBm
●
— — — — —
Max/Min
●
●
●
●
●
—
Relative
●
●
●
●
●
—
Hold
●
●
●
●
●
—
Compare
●
●
●
●
●
—
Math
●
●
●
●
●
—
Dual Measurement
●
●
●
●
— —
Common attribute: refresh rate
Background
Refresh rate defines how frequently GDM-8200 series
captures and updates the measurement data. Faster
refresh rate yields better accuracy but lower resolution.
Slower refresh rate yields lower accuracy but higher
resolution. Consider these trade-offs when selecting the
refresh rate.
Range S
5 ½ digits
M
4 ½ digits
F
3 ½ digits
Selection step
1.
Press the Shift key followed by
the AUTO (RATE) key. The
refresh rate switches to the next.
GDM-8200 Series User Manual
42
2.
The refresh rate indicator shows
the current status.
→ → →
Common attribute: reading indicator
Background
The reading indicator next to the 1st display flashes
according to the refresh rate when the captured data is
updated on the display.
When no data is
captured
When there is no captured data, the reading indicator
flashes once per two seconds (slower than the normal
refresh rate), showing GDM-8200 series is in the waiting
mode.
Common attribute: manual/automatic triggering
Automatic
triggering
(default)
GDM-8200 series triggers according to the refresh rate.
See the previous page for refresh rate setting details.
Manual
triggering
Press the TRIG key to trigger
measurement manually.