1.6.2 Measurement of the tooth length before restoration of post crown.
1.6.3 Measurement of the tooth length of transplantation and retransplantation.
1.6.4 The device must be operated in hospital and clinic by the qualified dentists.
1.7 Contraindication
We do not advise the use of the model on patients fitted with pacemakers (or other
electrical equipment) or on those patients who are advised not to use the electric
equipment (like electric shaver, electric blower) for safety reasons.
1.8 The classification of the device
1.8.1 Type of protection against electric shock: Class
Ⅱ
equipment
1.8.2 Degree of protection against electric shock: Type B applied part
1.8.3 Degree of protection against water shock: Ordinary equipment ( IPX0 )
1.8.4 Device not suitable for being used in the presence of a flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
1.8.5 Operation mode: Continous operation
1.8.6 Applied part: Touch probe, Lip hook, File clip
1.9 The main technical specifications
1.9.1 Battery: 3.6V/750mAh (Model:14500)
1.9.2 Adapter (Model:DJ-0500100-A5):
Input: ~100V-240V 50Hz/60Hz 0.5-0.2A
Output: DC5V/1A
1.9.3 Consumption power: ≤0.5W
1.9.4 Screen: 4.5’’ LCD