26
Operation
Operating mode
Warning
7. Inspection
- Inspect devices and sort out
those with defects.
- Dirty instruments must be cleaned and
disinfected again.
8. Packaging
- Pack the devices in
“Sterilization pouches”.
- Check the validity period of the pouch given
by the manufacturer to determine the shelf life.
- Use packaging which are resistant up to a
temperature of 141°C and in accordance with
EN ISO 11607.
9. Sterilization
- Steam sterilization at:
134
℃
, 2.0bar
~
2.3bar
(0.20MPa
~
0.23MPa), 4min.
- The accessories (lip hook, file clip and touch
probe) must be sterilized according to the
packaging labeling. - Use fractionated vacuum
or gravity (less preferred) autoclaves (according
to EN 13060, EN 285). - Use validated
sterilization procedure according to ISO 17665-
1 - Respect maintenance procedure of the
autoclave device given by the manufacturer. -
Use only the listed sterilization procedures.
10. Storage
- Keep devices in sterilization
packaging in a dry and clean
environment - Sterility cannot
be guaranteed if packaging is
open, damaged or wet (check
the packaging before using the
instruments).