2.6 Use a file size adapted to the root canal diameter. The selected file is too small
for a large root canal might cause the screen digital display is not steady during the
procedure.
2.7 In order to confirm the file clip and measuring wire makes good contact, test
the wire connecting before each use(See 3.1.3).
2.8 The file clip, lip hook and touch probe are reusable. Please make sure they
are autoclaved under high pressure and high temperature before each operation. The
endo files should not be used more than 3 times.
2.9 The batteries must be taken out for storage when the device is not used for a
long time.
2.10 Please recharge the battery when low battery indicator flashes.
2.11 Please use original components, the components made by other companies
may cause inaccurate measurement or un-measurable.
2.12 Avoid the connection between the outside and inside liquid of endodontic
during measuring in order to avoid the measuring difference.
2.13 Keep endo file and file clip away from any other metal or instruments.
2.14 To ensure that short circuits do not impair the measurements, be particularly
careful with patients fitted with metal crowns or bridges. Please confirm the wetness
of the endo to ensure the reliability of the measuring. If it is confirmed that the endo
file hasn't reached the apex yet the data showed on the apex locator is too low, please
check whether the endo is too dry and confirm it with X-ray.
2.15 This device have electromagnetic interference, the patient or doctor who