19
it is inconvenient to measure the back teeth [Picture 23].
Picture 23
4.2.7 Please remove the lip hook,file clip or touch probe after shutting down.
4.2.8 The components that touch body must be autoclaved under high temperature
and high pressure. The shell and measuring wire should be cleaned by 75% alcohol.
Attention:
Avoid the silk-screen when cleaning.
5 Trouble shooting
Problems
Possible causes
Solutions
No power and no signal on
the screen after the power on.
1. If the battery is placed
correctly?
2. If the battery with no
power?
1. Re-install the battery.
2. Recharge the battery.