8
will lead to the damage or the failure of the machine.
b) Measurement can not be proceeded without the complete insertion of the plug.
c) Be sure not to hit the plug. Keep the device away.
3.1.2 Insert the file clip and lip hook respectively into the two sockets of the
measuring wire. When the Apex Locator is used alone, there is no difference
between the gray end and the white end of measuring wire [Picture 3]. But if the
Apex Locator is connected to the Endo Motor, please connect the white end with
the lip hook and the gray end is suspended.
Attention:
Be sure not to pull the wire when inserting or pulling out the the measuring wire
and the file clip. [Picture 4 (a)]
Correct operation showed as in picture 4 (b).
3.1.3 Test the wire connecting (Test before each use)
a) Press the power switch. Make sure the scene of measuring the length of the
root canal displayed on the LCD screen.
The device will shutdown automatically after 5 minutes without operation.
b) Make sure if the plug of the measuring wire is inserted into the socket correctly.