CHAPTER 9    INTERFACE CONNECTION 

 

60 

 

9-4. ETHERNET CONNECTION 

(1)  Please use CAT 5 or higher grade cable. 

(2)  When connecting to the LAN with DHCP server and DNS, the DC power supply will obtain an IP automatically, 

this will take around 1 minute. 

(3)  The obtained IP can be read from the front panel, please refer to the 8-7-14~8-7-29 and 8-7-38~8-7-40 for detail, 

or it can also be read from the web browser, please refer to 9-4-3. 

(4)  Please refer to the Chapter 10. for the SCPI commands. 

(5)  Ethernet connection: 

 

9-4-1. Direct Connection 

 

 

9-4-2. Connection through HUB 

 

 

9-4-3. Through Internet or VPN 

 

 

 

Содержание Protek PD Series

Страница 1: ...INTRODUCTION 1 WWW GSI PROTEK NET User Manual Protek PR PD Series Programmable DC Power supply GSIPK ENP REV01 GS Instruments Co Ltd...

Страница 2: ...y the buyer Protek does not warrant supplied products or interfaces that are unauthorized modification misused operated outside of the specification repaired or maintained at the improper site No othe...

Страница 3: ...to mains supply voltage fluctuations 10 the maximum output power can be up to 750W at AC 100V input and decreasing linearity to 600W at 90V AC input LIVE CIRCUITS Operating personnel must not remove...

Страница 4: ...The WARNING sign denotes a hazard An attention to a procedure is called Not following procedure correctly could result in personal injury A WARNING sign should not be skipped and all indicated condit...

Страница 5: ...PTER 7 LOAD CONNECTION DESCRIPTION 31 7 1 WIRING INTRODUCTION 31 7 2 OUTPUT CONNECTION WITHOUT REMOTE SENSE 31 7 3 OUTPUT CONNECTION WITH REMOTE SENSE 32 7 4 CONSTANT VOLTAGE CONSTANT CURRENT CURVE 32...

Страница 6: ...ICATION 89 12 4 2UH 1500W SPECIFICATION 90 12 5 2U 3000W SPECIFICATION 90 APPENDIX 1 LOAD PROTECTION FUNCTION 92 A1 1 HIGH CAPACITANCE LOAD 92 A1 2 RECHARGEABLE BATTERY 92 A1 3 INDUCTIVE LOAD 92 A1 4...

Страница 7: ...ation manual 1 1 2 RS 485 Cable x 1 1 1 3 19 rack installation kit 1 set for 1U 2U models only 1 1 4 Output protective cover 6V 100V models only 1 1 5 Remote sense wire x2 Note 1 AC cables are not pro...

Страница 8: ...ng output Voltage Current Up to 5 units in parallel and 2 units in series equipped with active current sharing Active power factor correction Complete Protection OVP OCP and OTP Definable output volta...

Страница 9: ...ront Panel To maintain good accuracy please provide proper room for ventilation Do not block the ventilation holes in the front and the rear parts of DC power supplies Keep the panel with a space abov...

Страница 10: ...s LOCAL key to disable the remote control ALARM ALARM Indicator LED is lighting when there is a command error a manual operation error or power unit failure Please refer to the Chapter 11 for ERROR CO...

Страница 11: ...lue and then press ENTER to confirm 3 2 4 Output ON OFF indicator It s lighting when output is ON 3 2 5 Output ON OFF control key Output ON OFF enable disable The DC power supply output will be set to...

Страница 12: ...s V set key to adjust the voltage value and the last digit will be blinking By turning the Encoder voltage value is adjustable Press ENTER to confirm 2 SHIFT V set key perform as OVP set The process o...

Страница 13: ...TION 4 1 REAR PANEL 4 1 1 2U High Voltage Rear Panel 150V 600V models 4 1 2 2U Low Voltage Rear Panel 6V 100V models 4 1 3 1U High Voltage Rear Panel 150V 600V models 4 1 4 1U Low Voltage Rear Panel 6...

Страница 14: ...NEL DESCRIPTION 14 4 1 6 1UH Low Voltage Rear Panel 6V 100V models 4 1 7 2UH High Voltage Rear Panel 150V 600V models 4 1 8 2UH Low Voltage Rear Panel 6V 100V models 4 1 9 1UHx2 Rear Panel connection...

Страница 15: ...to avoid having interference it is recommended to connect a 1 4W 120 terminal resistor connect to and pin at the last unit 4 2 2 Digital programming Connector For GPIB remote control Optional interfa...

Страница 16: ...Parallel Operation for detail 4 2 4 When 2 units of PR PD same model are operated in series This connector will measure the output voltage of MASTER The SLAVE unit will adjust the output voltage to m...

Страница 17: ...rminals May be changed according to different capacities 4 2 6 1 1UH 2UH High Voltage 150V 600V models DC output terminals 4 2 6 2 1U 2U High Voltage 150V 600V models DC output terminals 4 2 6 3 1UH L...

Страница 18: ...CHAPTER 4 REAR PANEL DESCRIPTION 18 4 2 6 4 2UH Low Voltage 6V 100V models DC output terminals Side View Front View M6 Nut 2 pieces M6x15Hex Screw Washer 6 pieces...

Страница 19: ...CHAPTER 4 REAR PANEL DESCRIPTION 19 4 2 6 5 1U Low Voltage 6V 100V models DC output terminals Side View Front View M6 Nut 2 pieces M6x15Hex Screw Washer 2 pieces...

Страница 20: ...ls DC output terminals Side View Front View 4 2 7 AC INPUT terminal block The input terminals may be changed according to different capacities Note When 2U model s AC input underload error error code...

Страница 21: ......

Страница 22: ...l input 5Vdc for the relay of output ON OFF function 14 EXTV input common External voltage input V common Output ON OFF function 15 PRL IN For units those operate in parallel P15 is used for signal in...

Страница 23: ...on the rear panel to connect both of P21 and P25 To short or open P21 and P25 to control output ON OFF non isolated 2 Before controlling ON OFF by short open P21 P25 it is required to set EO C to be O...

Страница 24: ...open the P1 and P2 Please refer to 8 4 RECALL setting 5 1 8 OUTPUT STATUS MONITORING To connect P8 P9 P10 P11 P12 from APG on the rear panel APG Signal Name Description P8 State common Common for sta...

Страница 25: ...rtional output current measurement Use with Pin 20 Pin 9 10V Output Current Monitor of analog monitor 0 10Vdc for proportional output current measurement Use with Pin 21 Pin 10 GND Common ground Pin 1...

Страница 26: ...aces i e Isolated analog LXI GPIB 2 This option can only performs external voltage current control monitoring the other functions has to be controlled through the standard analog programming port plea...

Страница 27: ...1 Slave Connect Pin 1 of series in to MASTER output Connect Pin 2 of series in to MASTER output 2 If there is no series circuit board follow the wiring method below to connect analog programming ports...

Страница 28: ...age MASTER voltage SLAVE voltage j The current monitoring functions The APG connector Pin 22 A MON of MASTER and SLAVE are functional when these 2 units are used in series k When operated in series th...

Страница 29: ...the wiring method below to connect analog programming ports in each DC power 2 Please refer to Wiring introduction for the appropriate wire diameter According to the wiring diagram use the same length...

Страница 30: ...22 A MON of APG connector of MASTER and SLAVE are functional when the units are used in parallel The total output current MASTER current SLAVE current k When operated in parallel the protection functi...

Страница 31: ...5 15 2 6 1 1 8 8 8 0 2 97 5 48 8 24 4 9 8 3 0 6 14 0 1 152 4 6 1 38 1 15 2 4 9 4 22 0 1 243 8 121 9 61 0 24 4 7 9 2 38 0 1 365 7 182 9 91 4 38 1 12 2 0 60 0 0 609 6 304 8 152 4 61 0 20 7 Example When...

Страница 32: ...load side Please note that the compensation is varied in different models The above wiring method is recommended to the low noise loads 7 4 CONSTANT VOLTAGE CONSTANT CURRENT CURVE 7 4 1 Description o...

Страница 33: ...ecessary to accomplish this As follows 7 4 3 Description of Constant Voltage Constant Current CC CV curve A power supply acts as a constant voltage source for comparatively large values of load resist...

Страница 34: ...g on Executing the self test 8 1 2 Display the rated values voltage and current 8 1 3 Display the revision 1 1st window displays the Hardware revision 3 10 2nd window displays the Interface revision L...

Страница 35: ...be blinking Press ENTER to confirm The maximum adjustable value is 110 of rated voltage and the OVP setting on the panel will limit the adjustment no to be lower than the panel V setting Command SOUR...

Страница 36: ...ll remain at the last status 4 Press V set key to leave the RCL mode When the output is ON after pressing RCL the displays of voltage current windows will be blinking at the first 3 seconds When the o...

Страница 37: ...ce selection 8 6 9 GPIB 485 4 8K RS 485 baud rate setting 8 6 10 9 6K 19 2K 38 4K 57 6K 115 2K GPIB 0 30 GPIB address 8 6 11 485 001 254 RS 485 address 8 6 12 PSOP PM Parallel master 8 6 13 PS Paralle...

Страница 38: ...be only changed by pressing the ENTER key 1 DIRECT mode Displayed as 2 ENTER mode ENTER key must be pressed Displayed as 8 6 3 P ON POWER ON MODE When restarting the DC power recovering mains supply t...

Страница 39: ...are 16 memory sets in total MEM F is used to indicate the execute range of the set ones and the default value is set as 01 For example when the MEMORY FLAG is set as 05 it means the RECALL will only...

Страница 40: ...mode 1 RS 485 displayed as 2 GPIB displayed as 8 6 10 RS 485 Baud rate Baud rate setting for RS 485 the default setting is 115 2K 1 4800 displayed as 2 9600 displayed as 3 19200 displayed as 4 38400...

Страница 41: ...l or series operation maximum 5 units including MASTER for parallel or 2 units including MASTER for series PM Parallel Master PS Parallel Slave SM Series Master SS Series Slave 8 6 14 EO C External ou...

Страница 42: ...front panel EXT APG 8 6 17 CAL CODE CALIBRATION CODE An access to Calibration code Press ENTER key then turn the Encoder and enter the pin number 8 6 18 SN SERIAL NUMBER Indicate and display Serial nu...

Страница 43: ...time please consult us Note2 The ramp down time depends on different models For faster ramp down time please consult us 8 7 SETTING MENU FOR STANDARD LAN MODELS EQUIPPED WITH RS 485 LAN 1 Press SHIFT...

Страница 44: ...tion ON DHCP OFF DHCP ON DHCP ON DNS OFF DNS OFF DNS ON S IP1 V IP1 V IP1 001 223 Set View the 1 st IP address 8 7 14 S IP2 V IP2 V IP2 001 254 Set View the 2 nd IP address 8 7 15 S IP3 V IP3 V IP3 00...

Страница 45: ...8 7 32 EXT V set by APG CC V LOC A set by panel 8 7 33 EXT A set by APG CAL 0000 Access calibration mode 8 7 34 SN00 0000 Serial number total 6 numbers 8 7 35 FW x xx Firmware version 8 7 36 HW x xx H...

Страница 46: ...ins supply the output status is 3 LAST The output status will be sustained as the previous one that is before turning off the DC power or mains supply is interrupted Displayed as If the output status...

Страница 47: ...ECALL will only be executed from 01 05 8 7 7 BEEP KEY BEEP To set up the keying tone either to be turned ON or OFF Default setting of the keying beep sound is set as ON 8 7 8 BRIT DISPLAY BRIGHT Set t...

Страница 48: ...r RS 485 the default setting is 115 2K 7 4800 displayed as 8 9600 displayed as 9 19200 displayed as 10 38400 displayed as 11 57600 displayed as 12 115200 displayed as 8 7 11 485 RS 485 address RS 485...

Страница 49: ...front panel Refer to 8 7 14 8 7 17 or you may reset the IP to the default 192 168 0 100 Refer to 8 7 42 Set the acceptance of IP address assignment from DHCP server to ON OFF 1 Turn the Encoder to se...

Страница 50: ...mask gateway and DNS address correctly it is required to complete the setting of whole address i e S IP1 S IP4 before leave from the setting It indicates the 1 st IP address which assigned by the serv...

Страница 51: ...NTER key to access Turn the Encoder to change the number and then press enter key to confirm It indicates the 4th IP address which assigned by the server when DHCP ON The IP address has 4 numbers and...

Страница 52: ...255 For example the Subnet MASK address 255 255 xxx xxx This V MN3 shows the Subnet MASK address is 255 255 255 xxx 8 7 21 Set View the 4th Subnet MASK address The Subnet MASK address has 4 numbers an...

Страница 53: ...e Encoder to change the number and then press enter key to confirm It indicates the 3rd gateway address which assigned by the server when DHCP ON The gateway address has 4 numbers and each of them can...

Страница 54: ...n the Encoder to change the number and then press enter key to confirm It indicates the 2nd gateway address which assigned by the server when DHCP and DNS ON The DNS address has 4 numbers and each of...

Страница 55: ...rs and each of them can be set from 1 254 For example the gateway address 8 8 8 xxx This S DN4 function is to set the DNS address to 8 8 8 8 8 7 30 PSOP Parallel and serial operation Set the DC power...

Страница 56: ...nt Setting Selection The current setting adjustment can be selected to by using the front panel or APG LOC front panel EXT APG 8 7 34 CAL CODE CALIBRATION CODE An access to Calibration code Press ENTE...

Страница 57: ...1st and 2nd MAC number as 70 46 xx xx xx xx 8 7 39 To view the 3rd and 4th hex number of MAC address The MAC address contains 6 hex numbers i e 00 FF For example the MAC address 70 46 42 00 00 00 This...

Страница 58: ...to 0A 3 OVP is 110 of rated voltage 4 OCP is 110 of rated current 5 UVL is 0V 6 Output is OFF 8 7 42 RST1 IP address Reset Reset the IP address to 192 168 0 100 255 255 0 0 and clear the setting of G...

Страница 59: ...IB and RS 485 interfaces 6 RS 485 connection 9 2 USB CONNECTION 1 The DSP OPT USB convertor is available for option to control the DC power supply via USB interface 2 The DSP OPT USB convertor uses th...

Страница 60: ...IP automatically this will take around 1 minute 3 The obtained IP can be read from the front panel please refer to the 8 7 14 8 7 29 and 8 7 38 8 7 40 for detail or it can also be read from the web br...

Страница 61: ...Configuration Click the Network Configuration on the left side of the browser for the setting of IP configuration click Instrument Control for the control of the DC power supply and change the voltage...

Страница 62: ...9 4 6 Introduction of Network Status This page shows all the settings related to network 9 4 7 Log in screen The password is required when entering the Network Configuration and Instrument Control pa...

Страница 63: ...et Port when using the Hyper Terminal the default value is 5025 5 Loading the default settings 9 4 9 Introduction of instrument control There is a control screen for the basic controls of PR PD series...

Страница 64: ...h the exactly same content Execute the RST1 function in the MENU to load the default password if you forgot the new password please note all the settings those related to the Network will be reset to...

Страница 65: ...Terminal and key in the name of connection for example DSP HD 9 4 13 Select SOCKET PORT Select the Connect using to TCP IP Winsock and then key in the Host address and SOCKET PORT 9 4 14 click Proper...

Страница 66: ...CHAPTER 9 INTERFACE CONNECTION 66 9 4 15 ASCII Setup 1 Click ASCII Setup 9 4 16 ASCII Setup 2 Click Send line ends with line feeds and Echo typed characters locally...

Страница 67: ...CHAPTER 9 INTERFACE CONNECTION 67 9 4 17 Re connected again 9 4 18 Enter the commands for example IDN...

Страница 68: ...the A001 s display contrast to 3 5 Add a semicolom symble between two commands Example A001SYST REM A001DISP CONTrast 3 A001 SOURce VOLTage 30 6 When writing commands please use 0X0A as the end of lin...

Страница 69: ...ed voltage 10 and OCP is set to rated current 10 It is recommended that this command is always the first command set after remote digital control starts to ensure that the output is set to zero and co...

Страница 70: ...eset the LAN settings to the factory default state To enable or disable DHCP Query DHCP state To set a static Default Gateway Query gateway address To set a static DNS address Set DC power to auto obt...

Страница 71: ...the output voltage ramp up time Indicates how long output voltage takes to reach programmed value Sets the output voltage ramp dwon time Indicates how long output voltage takes to reach programmed va...

Страница 72: ...STem ERRor Description Returns error codes and messages Return Format string is text format in XXX XXXXXXXXXXXXX Example Input command S0URce VOLTage 2w will cause error SYSTem ERR returns error code...

Страница 73: ...ST BEEP Long Form SYSTem BEEP Description Indicates whether audible beep is on or off Returns 1 to indicate the beep is on 0 to indicate beep is off Return Format bool Example SYSTem BEEP returns 1 au...

Страница 74: ...text format in 70 46 42 xx xx xx Example SYST COMM LAN MAC Read the MAC address 70 46 42 xx xx xx is a typical default IP address 10 4 14 SYSTem COMMunicate LAN TELNet PORT COMMAND Syntax Short Form...

Страница 75: ...COMMunicate LAN DHCP Description Query DHCP state Return Format Bool Example SYST COMM LAN DHCP returns DHCP state 10 4 19 SYSTem COMMunicate LAN GATEway COMMAND Syntax Short Form SYST COMM LAN GATE a...

Страница 76: ...MMunicate LAN DNS Description Query DNS address Return Format address Example SYST COMM LAN DNS returns DNS address 10 4 24 SYSTem COMMunicate LAN DNS AUTO QUERY Syntax Short Form SYST COMM LAN DNS AU...

Страница 77: ...ON OFF Programs the unit to output off state upon power up or recovery from mains loss and allows OUTP command to set output to on or off 10 4 27 OUTPut PON QUERY Syntax Short Form OUTP PON Long Form...

Страница 78: ...cription Measure commands measure the output voltage Return Format Volts as a 5 digit number which is one decimal place more than the power supply s front panel voltmeter Example Example MEASure VOLTa...

Страница 79: ...ge protection OVP value in Volts Argument NRf is a value in the range from 0 to 110 of rated voltage Example SOUR VOLT PROT LEV 300 sets the overvoltage protection level to 300V 10 4 37 SOURce VOLTage...

Страница 80: ...RRent NRf MIN MAX Description Sets the output current value Argument NRf is a value in the range from 0 to rated current of unit Rated voltage and current are defined by model e g for DSP 030 025HD ra...

Страница 81: ...erating in constant current mode and exceeds 0 5 second Example SOUR CURR PROT STATE 1 enable the fold back protection 10 4 46 SOURce CURRent PROTection STATE QUERY Syntax Short Form SOUR CURR PROT ST...

Страница 82: ...CLS COMMAND Syntax Short Form SOUR MEM CLS Long Form SOURce MEMory CLS Description Clears all memory locations used for storing and recalling power supply settings Argument None Example SOURce MEMory...

Страница 83: ...n 5 to 62 4V 10 4 56 SOURce MEMory VOLTage NR1 QUERY Syntax Short Form SOUR MEM VOLT NR1 Long Form SOURce MEMory VOLTage NR1 Description Returns voltage value stored in memory location NR1 in Volts Re...

Страница 84: ...CV when operating in constant voltage mode 10 4 60 DSIP CONTrast COMMAND Syntax Short Form DSIP CONT NR1 Long Form DSIP CONTrast NR1 Description Sets the brightness of front panel Voltage and Current...

Страница 85: ...expression or string type was received that contained more data than the device could handle due to memory or related device specific requirements 224 Illegal parameter valid Used where exact value fr...

Страница 86: ...er 2 is failed including OTP of module 2 There are two DC to DC modules in 1500W models 77 Analog shut off shutdown a The slave unit s failed when operating in parallel series b The PIN23 and PIN24 of...

Страница 87: ...onal Isolated Voltage 1 Current 1 Temperature Coefficient 100ppm C from rated output after a 30 minute warm up Temperature drift 0 05 of rated output over 8 hours after a 30 minute warm up with consta...

Страница 88: ...ries models Humidity 30 90 relative humidity no condensation Storage environment Temperature 20 70 C Humidity 10 90 relative humidity no condensation EMC Standard EN 61326 1 2006 EMC Emissions EN 6132...

Страница 89: ...450 0 1 7 PR PD 4501 34 13 35 2 3 35 5 5 0 21 0 24 3 5 5 1UH 0 600 0 1 25 PR PD 6001 75 8 62 2 13 62 5 26 0 25 0 3 4 5 1UH 12 3 1U 1500W SPECIFICATION Output Model Ripple 3 Line regulation 5 Load regu...

Страница 90: ...0 0 3 4 PR PD 4503H 68 18 35 2 5 32 5 8 0 21 0 24 3 5 5 2UH 0 600 0 2 5 PR PD 6002H 113 15 62 2 26 62 5 5 0 25 0 3 4 5 2UH 12 5 2U 3000W SPECIFICATION Output Model Ripple 3 Line regulation 5 Load regu...

Страница 91: ...t voltage should be 0 1 of the rated voltage when the output voltage sets to zero 2 The actual output current should be 0 2 of the rated current when the output current sets to zero resistance load 3...

Страница 92: ...n the high capacitance load connects to the DC power supply please insert a diode in series to protect the DC power supply from OVP rapidly A1 2 RECHARGEABLE BATTERY Use the DC power supply in series...

Страница 93: ...ing cable ON and OFF simultaneously A1 5 WHEN THE LOAD DRAWS THE PULSE CURRENT When connected to the PWM load although some times the measured current value Mean or RMS is lower than the rated current...

Страница 94: ...will rise Please connect a resistor in parallel at the output stage to absorb the reverse current if the load regenerates a surge current please connect a capacitor in parallel GS instruments Co Ltd P...

Отзывы: