CHAPTER 8 OPERATION INTRODUCTION
35
8-2. VOLTAGE SETTING & OVP SETTING
Press V set key and turn the Encoder to adjust voltage value. The last digit will be blinking and adjustable.
(1) Enter the setting mode of output voltage (please refer to 8-6-2 OUT MODE in the MENU for setting methods).
Once you’ve finished the adjustment, press ENTER to confirm. The maximum adjustable value is 105% of rated
voltage, and OVP setting voltage will limit the adjustment not to be higher than the OVP value.
Command: SOURce:VOLTage
(2) Press SHIFT + V set key to make it perform as OVP setting. The value can be adjustable by turning the Encoder,
and the last digit will be blinking. Press ENTER to confirm. The maximum adjustable value is 110% of rated
voltage and the OVP setting on the panel will limit the adjustment no to be lower than the panel V setting.
Command: SOURce:VOLTage:PROTection:LEVel
(3) To avoid having restraint caused by commands, when changing operation from remote control to manual, it is
recommended to reset in MENU (refer to 8-6-21).
8-3. CURRENT SETTING & OCP SETTING
(1) Enter the output current setting mode (please refer to 8-6-2 OUT MODE in the MENU for settings). Press A set
key and turn the Encoder to adjust the current value. The last digit will be blinking. Press ENTER to confirm. The
A Set value will be limited by OCP value, not to be higher than the stored OCP value.
Command: SOURce:CURRent