Chapter 5: Preparing for Patient Use
Replace the Sensor Bands (if Using)
The sensor bands are designed to be used multiple times, but should be replaced between patients.
To place the motion sensor into a new strap, complete the following steps (Figure 12):
1. Remove the Kinesia Motion Sensors from the previous bands. We suggest discarding the old
straps as the Velcro can wear over time.
2. Select one wrist band (labeled with a hand) and one ankle band (labeled with a foot). Lay each
strap flat on the table with the buckle closest to you.
3. Take the Kinesia Motion Sensor for the upper extremity
(identifiable by the ‘U’ character beneath
the Sensor ID) and place it into the pocket on the wrist band. The Kinesia label should be facing
down and the LED indicator should be towards you.
4. Repeat the process above, placing the Kinesia Motion Sensor for the lower extremity
(identifiable by the ‘L’ character beneath the Sensor ID) into the ankle band.
5. Confirm that the Sensor ID of each sensor is visible.
Figure 14
Note:
To prevent the possibility of cross contamination and to maintain hygienic safety, a thorough
cleaning/disinfecting of the system components is required after each patient use. The sensor bands,
however, are single patient use only and should be discarded.