2-12
CHAPTER 2 Checks and Preparation
Alarm Output Specifications
Item
Description
Number of Output channels
4
Output 1 is used for alarm output and trigger output by switching.
Output format
Open collector output
+5 V, 10 kΩ pull-up resistance
* For details of alarm output, refer to the next page.
Trigger output time
When a start trigger or a stop trigger is detected, a pulse of approx.500 μs
width is output from the output 1 terminal.
(When setting low active and trigger output)
* When a stop trigger is issued within 1 ms from the start trigger, the trigger
is output at an interval of approx.1 ms.
When the power is turned OFF or ON, the device temporarily becomes at an alarm state .
Approx. 500μs
Approx. 1ms
Approx. 500μs
Trigger output
Trigger output
of start trigger
Trigger output
of start trigger