3-59
CHAPTER 3 Settings and Measurement
Available SPAN Settings
< Voltage Ranges >
Range
Maximum SPAN
Minimum SPAN
Minimum Resolution
20mV
-22.000 to +22.000mV
0.200mV
0.001mV
50mV
-55.00 to +55.00mV
0.50mV
0.01mV
100mV
-110.00 to +110.00mV
1.00mV
0.01mV
200mV
-220.00 to +220.00mV
2.00mV
0.01mV
500mV
-550.0 to +550.0mV
5.0mV
0.1mV
1V
-1.1000 to +1.1000V
0.0100V
0.0001V
2V
-2.2000 to +2.2000V
0.0200V
0.0001V
5V
-5.500 to +5.500V
0.050V
0.001V
10V
-11.000 to +11.000V
0.100V
0.001V
20V
-22.000 to +22.000V
0.200V
0.001V
50V
-55.00 to +55.00V
0.50V
0.01V
100V
-110.00 to +110.00V
1.00V
0.01V
200V
-220.00 to +220.00V
2.00V
0.01V
500V
-550.0 to +550.0V
5.0V
0.1V
1-5V
-5.500 to +5.500V
0.050V
0.001V
<Effective value (RMS) range>
Range
Maximum SPAN
Minimum SPAN
Minimum Resolution
10mVrms
-11.000 to +11.000mVrms
0.100mVrms
0.001mVrms
25mVrms
-27.500 to +27.500mVrms
0.250mVrms
0.002mVrms
50mVrms
-55.00 to +55.00mVrms
0.50mVrms
0.01mVrms
100mVrms
-110.00 to +110.00mVrms
1.00mVrms
0.01mVrms
250mVrms
-275.00 to +275.00mVrms
2.50mVrms
0.02mVrms
500mVrms
-550.0 to +550.0mVrms
5.0mVrms
0.1mVrms
1Vrms
-1.1000 to +1.1000Vrms
0.0100Vrms
0.0001Vrms
2.5Vrms
-2.7500 to +2.7500Vrms
0.0250Vrms
0.0002Vrms
5Vrms
-5.500 to +5.500Vrms
0.050Vrms
0.001Vrms
10Vrms
-11.000 to +11.000Vrms
0.100Vrms
0.001Vrms
25Vrms
-27.500 to +27.500Vrms
0.250Vrms
0.002Vrms
50Vrms
-55.00 to +55.00Vrms
0.50Vrms
0.01Vrms
100Vrms
-110.00 to +110.00Vrms
1.00Vrms
0.01Vrms
250Vrms
-275.00 to +275.00Vrms
2.50Vrms
0.02Vrms
< Temperature Ranges > : Celsius
Range
Maximum SPAN
Minimum SPAN
(p-p)
Measurable Range
Minimum
Resolution
K
-273.0 to +2000.0℃
10℃
-200 to +1370℃
0.1℃
J
-273.0 to +2000.0℃
10℃
-200 to +1100℃
T
-273.0 to +2000.0℃
10℃
-200 to +400℃
R
-273.0 to +2000.0℃
10℃
0 to +1600℃
E
-273.0 to +2000.0℃
10℃
-200 to +800℃
B
-273.0 to +2000.0℃
10℃
+400 to +1820℃
S
-273.0 to +2000.0℃
10℃
0 to +1760℃
N
-273.0 to +2000.0℃
10℃
-200 to +1300℃
W
-273.0 to +2000.0℃
10℃
0 to +2000℃