3-104
CHAPTER 3 Settings and Measurement
Logic trigger
The trigger of logic CH is compared with the pattern against setting the logic 4CH set in the mode.
Logic trigger
A trigger initiates when the input signal matches the set pattern.
Logic CH set to Off becomes “Don’t Care”, and it is regarded as a pattern match even if the input signal is H
or L.
However, when all logic CHs are set to Off, the logic trigger is invalid and is not used.
Specified time setting
By setting the time range of the specified time trigger, you can set the time trigger repeatedly.
Time setting range
Settable area
Description
Absolute time
Year, Month, Day, Hour,
Minute, Second
Specify the absolute time. The trigger is output only once.
Year
Month, Day, Hour, Minute,
Second
The trigger is output once a year.
Month
Day, Hour, Minute, Second
The trigger is output once a month.
Day
Hour, Minute, Second
The trigger is output once a day.
Hour
Minute, Second
The trigger is output once an hour.
Minute
Second
The trigger is output once a minute.
If the absolute time is set using the specified time trigger when trigger repeat setting is set to On, no trigger
occurs at repeat operation. (Trigger occurs only once.)
To set the repeat setting, be sure to set not using the absolute time.
Example:
Specified time range: hour
Time: 12 minutes 34 seconds
A trigger occurs every 12 minutes and 34 seconds.
1 hour 12 minutes 34 seconds
15 hours 12 minutes 34 seconds
A trigger occurs every hour as described above.