Micro-Ion E Module Instruction Manual 354011, Rev. A
43
Appendix A
Specifications
N
2
or Air Pressure Measurement Specifications
Measurements will change with different gases and mixtures.
Absolute Pressure Range
Torr
5 x 10
-7
to 5 x 10
–3
mbar
6.66 x 10
-7
to 6.66 x 10
-3
Pa
6.66 x 10
-5
to 6.66 x 10
–1
Overpressure Limit
Overpressure is the pressure at which the Micro-Ion E gauge shuts OFF with
increasing pressure.
Torr
1 x 10
–2
Pa
1.33 x 10
–2
mbar
1.33
X-ray Limit
X-ray limit is the absolute lowest indication from the gauge. It is not possible
to make repeatable measurements near the x-ray limit.
Torr
< 3 x 10
–10
mbar
< 4 x 10
–10
Pa
< 4 x 10
–8
Response Time
< 25 msec
Temperature Specifications
Operating Temperature
0 to 40 º C (32 to 104 º F) ambient, non-condensing
Storage Temperature
–40 to +70 º C (–40 to +158 º F)
8akeout Temperature
200 º C (392 º F) maximum with electronics removed