Series 580 TCD Gas Chromatograph
September 2017, Rev. 5
61
Symptom
Cause
Remedy
Round-topped peaks.
Operating beyond linear
Reduce sample size.
dynamic range of detector.
Recorder gain too low.
Adjust recorder gain. See
recorder manual.
Square-topped peaks.
Detector controller output
Normal for solvent peak.
exceeds recorder input
Reduce detector sensitivity,
range.
if desired.
Recorder slidewire
Check recorder operation.
defective or mechanism
See recorder manual.
binding.
Wrong attenuator setting.
Check attenuation.
Extra peaks.
Heavy residual material
Allow suffi cient time for
eluting from previous
previous sample injection to
sample injection.
elute.
(chromatogram I).
Chrom I
Air peak (chromatogram II).
Normal when making syringe
injections.
Desorption from column
Make several solvent
packing when solvent is
injections and recondition
injected (chromatogram II).
column.
Chrom II
Sample decomposition
Reduce injector temperature.
(chromatogram II).
Use diff erent column if
packing material is causing or
catalyzing decomposition.
Contaminated sample
Ensure proper preparation of
sample prior to injection.
Sample interaction with
Use diff erent column. Consult
mobile phase or column
GOW-MAC for column advise.
packing (chromatograms I,
II).
Contamination from glassware, Make sure all glassware,
syringes, etc. (chromatograms syringes, etc., are clean.
I, II).
Содержание 580 series
Страница 2: ...GOW MAC Instrument Co 2 ...
Страница 6: ...GOW MAC Instrument Co 6 ...
Страница 8: ...GOW MAC Instrument Co 8 ...
Страница 10: ...GOW MAC Instrument Co 10 ...
Страница 24: ...GOW MAC Instrument Co 24 ...
Страница 36: ...GOW MAC Instrument Co 36 ...
Страница 42: ...GOW MAC Instrument Co 42 ...
Страница 54: ...GOW MAC Instrument Co 54 ...
Страница 62: ...GOW MAC Instrument Co 62 ...
Страница 66: ...GOW MAC Instrument Co 66 ...
Страница 68: ...GOW MAC Instrument Co 68 ...