Series 580 TCD Gas Chromatograph
September 2017, Rev. 5
45
Figure 7-1
Column Placement
3. Detector Removal
CAUTION
BEFORE ATTEMPTING TO REMOVE THE DETECTOR OR
FILAMENTS, ALL POWER TO THE INSTRUMENT MUST BE
TURNED ”OFF” BY REMOVING THE POWER CORD FROM THE
OUTLET.
CAUTION
INSULATION CONTAINS FIBERGLASS. USE OF FACE MASK &
GLOVES IS RECOMMENDED.
To remove the detector, use the drawings provided and proceed as follows:
a. Remove the columns using 9/16” open-ended wrenches for 1/4” fi ttings or 7/16”
wrenches for 1/8” fi ttings.
b. Remove the two (2) screws located just above and below the OUTLET PORT .
c. Remove the SEPTUM NUTS.
d. Remove the six (6) screws from the FRONT OVEN PANEL.
e. CAREFULLY remove insulation so as to expose the DETECTOR HOUSING.
f. Remove the four (4) screws located directly above the lower column connections inside
the column oven.