Series 580 TCD Gas Chromatograph
September 2017, Rev. 5
29
Section 5
Operation
A. General
The chromatographer should be familiar with the techniques of chromatography, the function of
all controls, the operation of the recorder, and the characteristics of the columns used prior to
running samples.
B. Operation
1. Gas Flow
a. Make sure that ALL switches are in the “OFF” position.
b. Set helium or other carrier gas pressure regulator to 40 psig.
c. Adjust fl ow of carrier gas to: 30 mL/min. for 1/8” columns 60 mL/min. for 1/4” columns.
d. Check for leaks as described in Section 3, Paragraph I.
e. Disconnect fl owmeter.
f. Allow 5 minutes to purge the system before turning power “ON”.
2. Temperature Controls
NOTE
)
TEMPERATURE CONTROL KNOBS are locking type. Push locking
ring “IN” to turn knob.
a. Turn the DETECTOR CURRENT CONTROL fully counterclockwise (CCW).
THIS STEP
IS EXTREMELY IMPORTANT!
b. Plug in the instrument to the appropriate ac outlet. Switch the instrument “ON”.
Charts 5-1 thru 5-5 illustrate the maximum current for specifi c cell temperature
and carrier gases helium, nitrogen, and argon for single and dual carrier detectors.
THESE SHOULD NOT BE EXCEEDED!
However, for longer fi lament life and a more
stable baseline, lower detector current is recommended. Reference recommended
starting bridge settings (pg. 26).
Содержание 580 series
Страница 2: ...GOW MAC Instrument Co 2 ...
Страница 6: ...GOW MAC Instrument Co 6 ...
Страница 8: ...GOW MAC Instrument Co 8 ...
Страница 10: ...GOW MAC Instrument Co 10 ...
Страница 24: ...GOW MAC Instrument Co 24 ...
Страница 36: ...GOW MAC Instrument Co 36 ...
Страница 42: ...GOW MAC Instrument Co 42 ...
Страница 54: ...GOW MAC Instrument Co 54 ...
Страница 62: ...GOW MAC Instrument Co 62 ...
Страница 66: ...GOW MAC Instrument Co 66 ...
Страница 68: ...GOW MAC Instrument Co 68 ...