GOW-MAC Instrument Co.
48
3. Changing Columns
Figure 7-2 shows the injection port for low dead volume and on column injection. Either 1/4”
or 1/8” o.d. columns can be used as shown. All 1/8” o.d. columns are furnished with 1/8”
nuts and ferrules.
Care should be exercised when changing or removing columns. Damage to adjacent threads
may occur if they are hit with a wrench or other object which may result in nuts becoming
cross threaded.
Care should also be used when inserting the columns into the injection port assembly. Insert
the column until it stops, making sure that it has reached the bevelled end of the injection
port.
Then back off about 1/4” to allow carrier gas to sweep into the column. By using this
technique, on- column injections can be attained. CAREFULLY tighten nuts with a 9/16”
wrench (or 7/16” wrench for 1/8” fi ttings).
CHECK FOR LEAKS.
1/4" o.d. Column
1/8" o.d. Column
with Glass Adapter
1/8" o.d. Column
(Standard)
(Optional)
(Optional)
Figure 7-2
Injection Ports
C. Septums
The septums used in this instrument are standard 9 mm o.d. and may be obtained directly from
GOW-MAC (Part No. 180-123). The INJECTION PORT NUT may be removed and the SEPTUM
replaced. POWER TO THE GC MUST BE “OFF” and the instrument cool BEFORE removing the
INJECTION PORT NUTS. The INJECTION PORT NUT acts as a heat sink, and as such, should
be kept clean and polished.
Air Space
Air Space
Air Space
Содержание 580 series
Страница 2: ...GOW MAC Instrument Co 2 ...
Страница 6: ...GOW MAC Instrument Co 6 ...
Страница 8: ...GOW MAC Instrument Co 8 ...
Страница 10: ...GOW MAC Instrument Co 10 ...
Страница 24: ...GOW MAC Instrument Co 24 ...
Страница 36: ...GOW MAC Instrument Co 36 ...
Страница 42: ...GOW MAC Instrument Co 42 ...
Страница 54: ...GOW MAC Instrument Co 54 ...
Страница 62: ...GOW MAC Instrument Co 62 ...
Страница 66: ...GOW MAC Instrument Co 66 ...
Страница 68: ...GOW MAC Instrument Co 68 ...