Series 580 TCD Gas Chromatograph
September 2017, Rev. 5
13
Section 1
Specifi cations
(dependent upon the number of options ordered)
OVERALL
:
H 12 1/2”
(317 mm)
W 19 1/2”
(495 mm) width varies due to housing used for options ordered
D 18”
(457 mm)
Net Weight: 70 lbs. (31.75 kg)
Shipping: 80 lbs. (36.21 kg)
POWER REQUIREMENTS:
Series 580: 105 - 125 Vac, 50/60 Hz, 1100 W
Series 582: 200 - 240 Vac, 50 Hz, 1100 W
Fuse: Series 580: 10 Amps
Series 582: 5 Amps
COLUMN OVEN:
H 7 1/2” (190 mm)
W 10” (254 mm)
D 8 1/2” (216 mm)
Temperature Range: ambient to 400 °C
Temperature Readout: 3 1/2 digit LED digital meter Temperature Control: solid state, time
proportioning, RTD sensors, direct reading, ambient to 400 °C
Column Oven Temperature Protection Circuit: shuts the column oven off if the temperature
rises to 30 °C over the set point. Oven Fittings: accommodates 1/8” or 1/4” o.d. metal,
6 mm glass, or capillary columns.
Oven Capacity: can accommodate up to two (2) 1/4” o.d. x 20’ columns or correspondingly
longer lengths of 1/8” o.d.
DETECTOR OVEN:
Temperature Settings: ambient to 400 °C
Temperature Readout: 3 1/2 digit LED digital meter
Temperature Control: solid state time proportioning, RTD sensors, direct reading, ambient
to 400 °C
Содержание 580 series
Страница 2: ...GOW MAC Instrument Co 2 ...
Страница 6: ...GOW MAC Instrument Co 6 ...
Страница 8: ...GOW MAC Instrument Co 8 ...
Страница 10: ...GOW MAC Instrument Co 10 ...
Страница 24: ...GOW MAC Instrument Co 24 ...
Страница 36: ...GOW MAC Instrument Co 36 ...
Страница 42: ...GOW MAC Instrument Co 42 ...
Страница 54: ...GOW MAC Instrument Co 54 ...
Страница 62: ...GOW MAC Instrument Co 62 ...
Страница 66: ...GOW MAC Instrument Co 66 ...
Страница 68: ...GOW MAC Instrument Co 68 ...