Software
English, Revision 02, Date: 12.07.2011
41
5.2.2.4
P(a)rallel Interface
Figure 17
Menu P(a)rallel Interface
In this menu the characteristics of the parallel interface can be set. It is only important
for
interpreter variant Y.
The duration of the data ready pulse DTR (
) can be modified using 1 ms steps. With
DTR Seup Time (
) the time from creating the parallel code to the output of the DTR
code can be determinded (see also appendix, chapter F "Timing-Diagramm of the
parallel interface" on page 57).
With
you can determine whether only one transponder with a new code generates
a DTR pulse (1) or if the same transponder can generate two pulses by crossing the
detection area and leaving it again (0).
With
all outputs can be set to 1 or 0 for test purposes. The outputs DTR and Po-
sipuls will be active for the set time. After leaving the monitor mode all outputs will be
turned off again.
S:0020 D:+0022 Code: 000fffff Read: 114 E: 0120 N: 0 Ftx[Hz]:128010
DT(R)-Pulse Time [n*1ms]: 100
DTR-(S)etup Time [n*1ms]: 50
(1) DTR - Pulse per Code 0
(O)uttest
(Q)uit Menue