GOSSEN-METRAWATT
7
1.7 Switching the recorder on
Caution
Prior to switching the power supply on, verify that the operating voltage
of the recorder (see nameplate) and the supply voltage agree.
A power line connection switch of sufficient switching capacity,
which permits all-pole disconnection of the recorder from the
power line, must be provided within reach of the mounting site. It
must not annul the protective effect of the protective conductor.
Date and time are buffered via a capacitor (Super Cap). After
longer periods without auxiliary voltage (e.g. when in stock), date
and time are reset to a default value. At the next start, a note
appears on the display (error message E150X). Date and time
must then be set newly (see parameterizing instructions, ordering
number 14083).
1.8 Positioning the recording chart
(see figure 10)
1. Press the lower grip boards of the recording table to the rear.
The recording chart is transported with increased speed in the
direction of flow.
2. Release the grip boards when the desired time line is reached.
Figure 10 Positioning the recording chart
2 Operation
2.1 Removing the recording chart
(see figure 11)