10
GOSSEN-METRAWATT
Figure 17 Arrangement of jumpers as a relation of measuring mode and
nominal measuring range
3.2 Replacing scales
(see figure 18)
Figure 18 Replacing scales
1. Withdraw the fiber pen inserts.
2. Undo the scale screws at left.
3. Slide the scales to the right and disengage them from the scale
screw.
4. Remove the scales to the left.
5. Install the scales in reverse order.
6. Install the fiber pen inserts.
7. Check the measuring system zero with beginning of scale.
a. Remove the recording table.
b. Press <
↵
>. "SYS" is displayed.
The measuring systems travel to electrical zero.
c. Align scale with pointer and tighten the scale screw.
d. Insert the recording table.
3.3 Replacing the label for the measuring points
(see figure 19)
Figure 19 Replacing the label for the measuring points
• Remove the flexible label for the measuring points from its
holder and replace it with a new one.
1 2
1 2
1 2
1 2
1 2
1 2
1 2
1 2
Current
0...20 mA
4...20 mA
Current
–20...20 mA
Thermocouple
Type B,R
S,T
Type E,J, K
L,N, U
Voltage
–75...75 mV
Voltage
–20...20 mV
Pt 100
–50...150
°
C
Pt 100
–50...500
°
C
Jumper plugged-in