If there is a need to compare the waveforms, the method of channel
order setting (see Channel Order Setting) can be used to display the
desired waveforms.
Waveform Rolling Left/Right
The Model 3600 allows for 260,000 memory addresses for each
channel, however only 280 data points can be viewed due to the
horizontal width of the screen. In order to display the needed blocks
from the mass of stored data, the screen display window needs to be
repositioned to the desired data in memory. This requires setting a
changeable “window address” for the display window. The screen of
the analyzer will show the stored data block using this new window
address as starting point.
The parameter in the first row on the right bottom of the wave frame
is the window-address. The waveform displays the data on the most
left, that is the data corresponding to the window address in sampling
memory. Press
【
System
】
to select window-address and set the window
address in Dec numbers with the range of 0~260000. After setting,
the waveform of sampled stored data block associated with this new
window address will be displayed. To continuously observe a waveform
on the display, press
【
← →
】
and turn the knob. The window-address
will change continuously showing the waveform rolling from left to
right in the display window.
The parameter in the second row on the right bottom of the wave
frame is a scroll-step function. Turn the knob one step and the window-
address will increase one scroll-step value. The higher the value of
scroll-step, the faster the speed of the rolling waveform. This could
possibly surpass the desired data block. The lower the value of scroll-
step, the slower the speed of rolling and the finer observation of the
waveforms. Press
【
System
】
to select scroll-step. The value of the scroll-
step can be set by using the number keys with the range of 1~ 260 000.
After setting, press
【
← →
】
and turn the knob to roll the waveform
left and right, so that the speed of rolling waveform will change.
Setting the window-address using the number keys directly enables
display window to locate the desired sampled block of data. Rolling
waveforms utilizing the knob allows observation of section changes in
a continuous manner.