gradually to minimize distortion. If distortion is found, the user should
no longer continue compressing the waveform.
Data Lookup - Search Function
Model 3600 has a convenient search function. After a sampling of
data, the user can find the data conforming to the setting conditions
within a large group of sampled data. Parameters in the third row on
the right bottom of the wave frame is the find-data which is a 32 bit
data word in Hex format. Press
【
System
】
to select find-data function.
Select search data word and input the numbers from 0~9 or the letters
of A~F. User can also input x (Note: the
【
x
】
on the right of the key
【
0
】
is not the letter “x” in English). x means “ignore”. After setting the
find data word, press
【
Find
】
again to locate the data according to the
search conditions. This will be denoted with a yellow dash, generally
seen on the left of the screen. At the same time, the window address
will change to show the position of this data in the sampled memory.
Press
【
Find
】
repeatedly until reaching the end point of sampled data
memory. This allows user to find all the data words conforming to
setting conditions from sampled data.
The find-data function runs after finishing the data sampling. It is used
to look up a data word set randomly in stored sampling data. This
point is different from the trigger conditions data word that will be
described in another section. Trigger conditions is preset in advance
before sampling and used to capture data word conforming to trigger
conditions.
Data Listing Display
Press the
【
Display
】
key and 19 rows of data listings will be displayed.
In the data listing interface, the most left line is the Dec address
value of sampled data in memory. The middle is the address value
of the sampled data in Hex, the right is the Bin address value. A
differentiating line is displayed between each 8 bits Bin code for
convenient reading. There are two different colors to distinguish the
two adjacent lines to allow for clear viewing and eliminate confusion.
The storage depth of the instrument is 260,000 storing addresses, but
the screen shows only 19 rows in a vertical manner. Therefore, data
lists also must roll up and down in the display window. The parameter
definitions and operation methods shown in the data listing display