PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
PRO-series Model E3 Electrodeless Conductivity Transmitter
7
5.8 RESET CALIBRATE
Values to Factory
Defaults
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display the “RESET CONFIGURE:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (To abort this procedure,
press
ESC key
now.)
3.
Press
ENTER key
to reset stored configuration settings
--
not calibration settings
-- to factory defaults. The
“RESET CONFIGURE: DONE” screen appears, ac-
knowledging that reset has occurred.
4.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.
The RESET CALIBRATE function resets all stored calibra-
tion settings --
but not configuration settings
-- to factory-
set defaults.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display the “RESET CALIBRATE:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (To abort this procedure,
press
ESC key
now.)
3.
Press
ENTER key
to reset all stored calibration settings
--
not configuration settings
-- to factory defaults. The
“RESET CALIBRATE: DONE” screen appears, ac-
knowledging that reset has occurred.
4.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.