PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
PRO-series Model E3 Electrodeless Conductivity Transmitter
☞
5.4 Firmware (EPROM
VERSION) Check
5.5 SELECT SIM
Measurement
2.
Press
ENTER key
to display a screen like
.
NOTE:
The mA output test signal is now active. Its
value is shown on this screen.
3.
Use
arrow keys
to adjust the displayed value to obtain
the desired mA test signal.
4.
To remove the output test signal and return to the pre-
vious level of the TEST/MAINT menu branch, press
ESC key
or
ENTER key
.
The EPROM VERSION function checks the version of firm-
ware used in the transmitter.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to view the EPROM version screen.
3.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.
The SELECT SIM function selects a type of simulated
measurement. It is used in conjunction with the SIM
SENSOR function (Section 5.6) to simulate a measured
value, making the analog output respond accordingly.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to select the
type of simulated measurement, and press
ENTER key
to enter it: