PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
PRO-series Model E3 Electrodeless Conductivity Transmitter
5.6 SIM SENSOR Setting
☞
5.7 RESET CONFIGURE
Values to Factory
Defaults
☞
•
SENSOR:
Selects simulated measurement to be
the configured measurement (conductivity, % con-
centration or TDS).
•
TEMPERATURE:
Selects simulated measurement
to be temperature.
After selecting the type of simulated measurement (Section
5.5), use the SIM SENSOR function to set the desired
simulation value.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
.
NOTE:
The analog output signal is now active. It has a
mA value that corresponds to the measurement
value shown on this screen.
3.
Use
arrow keys
to adjust the displayed simulation
value to the desired value.
4.
To remove the simulated output and return to the previ-
ous level of the TEST/MAINT menu branch, press
ESC
key
or
ENTER key
.
The RESET CONFIGURE function resets stored configura-
tion settings (all at the same time) --
but not calibration
settings
-- to their factory-set defaults shown in TABLE D.
NOTE:
Resetting configuration values also excludes the
SELECT MEASURE function (conductivity, % con-
centration or TDS) which remains as is until you
change it.