PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
3
☞
CONFIG CONC or
CONFIG TDS Measurement
(configuration not needed
for conductivity)
•
NATURAL WATER (not available for TDS meas-
urement):
Built-in temperature properties table only
for special applications -- consult factory)
•
TEMP TABLE:
User-defined temperature table
•
NONE:
Measurement values are not compensated
NOTE:
LINEAR is the factory default for temperature
compensation with a 2.00% per
°
C slope and
25.0
°
C reference temperature. This provides
the best results for most aqueous solutions. To
enter different slope and reference temperature
values for an uncommon solution, refer to sub-
heading “CONFIG LINEAR or CONFIG T-TABLE
Temperature Compensation” for details.
Only when CONCENTRATION or TDS is selected must
the transmitter be further configured.
If CONDUCTIVITY
was selected, disregard this subsection -- no measurement
configuration is needed.
CONCENTRATION Measurement Setup
Configure the transmitter with an appropriate table to con-
vert measured conductivity into displayed % concentration.
If one of the transmitter’s BUILT-IN chemical concentration
tables matches the solution being measured, simply select
that table. If not, you must create a USER-DEFINED con-
centration table for the solution being measured.
Selecting BUILT-IN Chemical Concentration Table
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display
.
BUILT-IN configures transmitter to use one of the built-
in chemical concentration tables. (If screen shows USER-
DEFINED, use
Ø
Ø
and
×
×
keys
to select BUILT-IN.)
4.
With “BUILT-IN” displayed, press
ENTER key
.