PART THREE - OPERATION
SECTION 4 - TRANSMITTER CALIBRATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
SAMPLE CAL Method
☞
The “SAMPLE CAL” method enables the sensor to remain
installed in the process.
1.
Obtain a sample of the process solution and determine
its value using laboratory analysis or a recently cali-
brated portable meter.
2.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
3.
Press
ENTER key
to display
.
4.
Press
ENTER key
again to display
.
5.
Press
Ø
Ø
key
once to display
.
6.
Press
ENTER key
to display
.
With the sensor in the process, press
ENTER key
again to confirm. This active
screen appears showing the measurement reading.
NOTE:
During calibration, the analog output is auto-
matically “held” at the last measured value.
7.
Wait for the reading to stabilize which may take up to
30 minutes. Then press
ENTER key
. The “PLEASE
WAIT” screen may appear if the reading is still too un-
stable. After the reading has stabilized, this static
screen appears showing the “last
measured” value.
8.
Use
arrow keys
to adjust the displayed value to
exactly match the known value of the process sample.
9.
Press
ENTER key
to enter the value and complete cali-
bration (“CONFIRM CAL OK?” screen appears).
10. Press
ENTER key
again to display the active meas-
urement reading on the “CONFIRM ACTIVE?” output
status screen. When the reading corresponds to the