PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
3
☞
Select
DISPLAY FORMAT
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display a screen like
. Use
Ø
Ø
and
×
×
keys
to select the
desired measurement (conductivity, concentration or
TDS), and press
ENTER key
to enter it.
NOTE:
If concentration was selected, measured con-
ductivity must be converted to % concentration
by selecting a BUILT-IN chemical concentration
table or creating a USER-DEFINED table. See
“CONFIG CONC” subheading for details.
WARNING:
CHANGING THE MEASUREMENT AUTOMATICALLY
REPLACES ALL USER-ENTERED CONFIGURA-
TION VALUES WITH FACTORY-DEFAULTS.
After choosing the measurement, select the desired display
format for the MEASURE screen. The selected units and
resolution will also appear on all applicable edit/selection
menu screens.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Refer to the selected measurement category below and
follow its steps:
CONDUCTIVITY Display Format
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to select a
format (2000 μS/cm, 200.0 μS/cm, 2.000 mS/cm,
20.00 mS/cm, 200.0 mS/cm, 2000 mS/cm, or 2.000
S/cm), and press
ENTER key
to enter it.