PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
4
CONFIG LINEAR or
CONFIG T-TABLE
Temperature Compensation
(configuration not needed for
other compensation methods)
Only when LINEAR or TEMP TABLE is the selected
temperature compensation, must the transmitter be
further configured.
If the built-in NATURAL WATER prop-
erties table or NONE was selected, disregard this
subsection -- no compensation configuration is needed.
LINEAR Compensation Setup
Factory defaults for LINEAR compensation are 2.00%/
°
C
slope and 25.0
°
C reference temperature.
These values are
appropriate for most aqueous solutions.
Use chemical
handbook tables to find values for uncommon solutions. To
enter different values:
1.
With the
or
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display a screen like
. Use
arrow keys
to adjust to a
desired slope, and press
ENTER key
to enter it.
4.
After the
screen re-appears, press
Ø
Ø
key
once to display
.
5.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust to a
desired reference temperature, and press
ENTER key
to enter it.
6.
After the
screen re-appears, press
ESC key
once to return to the
screen.