PART THREE - OPERATION
SECTION 8 - HART OPTION
Rev. 6-501
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
83
8.4 MULTI-DROP
Wiring Arrangement
☞
When the GLI analyzer is set to operate in the MULTI-
DROP mode on a HART network, the master(s) is intended
to communicate with multiple analyzers.
NOTE:
When analyzers are operated in the MULTI-DROP
mode, the 4-20 mA analog OUTPUT 1 signal of
each analyzer is dedicated only for network use --
not its normal use. (During startup, each analyzer is
assigned a non-zero polling address, causing its
Output 1 to automatically provide a constant 4 mA
signal.) Each analyzer’s analog OUTPUT 2, how-
ever, remains available for normal use.
1.
Make sure the
SINGLE MODE/MULTI-DROP
switch of
each analyzer is set to the
MD
(right) position.
2.
Refer to Figure 3-4 and connect the 4-20 mA analog
Output 1 signal of each analyzer in parallel on one
cable, matching polarity as shown.
3.
Connect an appropriate-sized power supply in parallel
with the analog Output 1 signal, matching polarity as
shown.
4.
Up to two masters can be connected to the 4-20 mA
analog Output 1 signal cable.
FIGURE 3-4 HART MULTI-DROP Wiring Arrangement (for multiple analyzer network)