PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Rev. 6-501
Model P53 pH/ORP Analyzer (universal-mount 1/2 DIN)
77
6.10 RESET CAL
Values to Factory
Defaults
1.
With the
screen displayed, use
Ø
Ø
key
to select the “RESET CONFIG” line.
2.
Press
ENTER key
to display the “RESET CONFIG:
ARE YOU SURE?” screen, asking if you really intend to
perform this extreme action. (To abort this procedure,
press
ESC key
now.)
3.
Press
ENTER key
to reset all stored configuration set-
tings --
not calibration settings
-- to factory defaults.
The “RESET CONFIG: DONE” screen appears, ac-
knowledging that reset has occurred.
4.
To return to the TEST/MAINT top-level menu screen,
press
ESC key
or
ENTER key
.
You can conveniently reset all stored calibration settings --
but not configuration settings
-- to factory-set defaults.
1.
With the
screen displayed, use
Ø
Ø
key
to select the “RESET CAL” line.
2.
Press
ENTER key
to display the “RESET CAL: ARE
YOU SURE?” screen, asking if you really intend to
perform this extreme action. (To abort this procedure,
press
ESC key
now.)
3.
Press
ENTER key
to reset all stored calibration settings
--
not configuration settings
-- to factory defaults. The
“RESET CAL: DONE” screen appears, acknowledging
that reset has occurred.
4.
To return to the “TEST/MAINT” top-level menu screen,
press
ESC key
or
ENTER key
.